PURPOSE AND SCOPE

Gas analyzer DAG - 500 for measuring the concentration of carbon monoxide, oxygen, sulfur dioxide, nitrogen oxide, nitrogen dioxide, temperature measurement, pressure / vacuum and soot emissions in gas fuel combustion plants.

DESCRIPTION

1. Gas analyzer DAG - 500 is a portable automatic multichannel showing a continuous apparatus.

2. The measuring part analyzer DAG - 500 used method for analysis of multicomponent gas mixtures through a set of electrochemical sensors having selective sensitivity to the various components of the gas mixture and producing electrical signals proportional to the measured concentration of components.

3. The flue gas temperature is measured using a thermocouple XA (K) placed in the controlled area.

4. Structurally analyzer DAG - 500 is a rectangular shape measuring unit, the front panel is a keypad, a display. On the front panel of the analyzer are:

  • a connector for external power supply " + 12";
  • connector for connection of the probe with a thermocouple;
  • connector for connecting a cable for connection to a PC or an external printer;
  • a connector for connection to an external sensor cable;
  • fitting for sampling the gas sample;
  • fitting for pressure / vacuum.

5. analyzer DAG - 500 is designed for operation in explosion - proof environment in a continuous, intermittent, re - intermittent modes.

6. The measurement results (in% in mg / m 3 , ppm) can be recorded in nonvolatile memory, output to an external printer or the IBM PC type computer through the RS232 interface.

7. analyzer DAG - 500 has the ability to connect external sensors.

 

 

TECHNICAL SPECIFICATIONS

Gas analyzer DAG - 500 provides a measurement of the flue gas parameters with characteristics in accordance with Table. one.

Table 1.

Identify the components range of  measurement Output accuracy The nominal price of a unit of the smallest discharge
absolute relative
Oxygen (O 2 ) 0 - 21% vol. ± 0, 2% vol. - 0, 1% vol.
Carbon monoxide (CO) 0 - 30 000 ppm ± 300 ppm (0 - 6000 ppm) ± 5% (6, 000 - 30, 000 ppm) 1 ppm
0 - 6000 ppm ± 60 ppm (0 - 1200 ppm) ± 5% (1200 - 6000 ppm) 1 ppm
0 - 2000 ppm ± 20 ppm (0 - 400 ppm) ± 5% (400 - 2000 ppm) 1 ppm
0 - 1000 ppm ± 10 ppm (0 - 200 ppm) ± 5% (200 - 1000 ppm) 1 ppm
0 - 500 ppm ± 5 ppm (0 - 100 ppm) ± 5% (100 - 500 ppm) 1 ppm
Nitric oxide (NO) 0 - 2000 ppm ± 40 ppm (0 - 400 ppm) ± 10% (400 - 2000 ppm) 1 ppm
0 - 1000 ppm ± 20 ppm (0 - 200 ppm) ± 10% (200 - 1000 ppm) 1 ppm
0 - 500 ppm ± 10 ppm (0 - 100 ppm) ± 10% (100 - 500 ppm) 1 ppm
0 - 250 ppm ± 5 ppm (0 - 50 ppm) ± 10% (50 - 250 ppm)

1 ppm

Nitrogen dioxide (NO 2 ) 0 - 100 ppm ± 3 ppm (0 - 20 ppm) ± 15% (20 - 100 ppm) 1 ppm
0 - 50 ppm ± 2 ppm (0 - 10 ppm) ± 15% (10 - 5 ppm) 1 ppm
Sulfur dioxide (SO 2 ) 0 - 4000 ppm ± 80 ppm (0 - 800 ppm) ± 10% (800 - 4000 ppm) 1 ppm
0 - 2000 ppm ± 40 ppm (0 - 400 ppm) ± 10% (400 - 2000 ppm) 1 ppm
0 - 1000 ppm ± 20 ppm (0 - 200 ppm) ± 10% (200 - 1000 ppm) 1 ppm
0 - 500 ppm ± 10 ppm (0 - 100 ppm) ± 10% (100 - 500 ppm) 1 ppm
The temperature of the gas stream - 20 - + 800 on the C ± 3 on C ( - 20 - 300 + of C) ± 1% (300 - 800 of C) 1 of C
ambient temperature 0 - 60 of the C ± 1 for C - 1 of C
Pressure / vacuum ± (0 - 5, 0) kPa 0 ± 02 kPa (0. . . 0, 4 kPa) ± 5% (0, 4. . 5 kPa) 1 Pa
Carbon dioxide (CO 2 ) not normalized (the definition of convenience)

 

  • The limit of permissible variation indications, b d is 0, 5, the maximum permissible error.
  • Warm - up time, with no more than 240
  • Setting time, with no more than 180
  • The interval time of the uncorrected readings hours 1000
  • Supply voltage, V (220)
  • Power consumption is not more than 10
  • Frequency, Hz (50 ± 1)
  • Mean time between failures, h, not less than 10, 000
  • Average service life, years, not less than 8
  • Overall dimensions, mm, not less 220h105h65
  • Weight, not more than 1, 5

Terms of Use:

  • operating temperature range of C 10 - 40
  • relative humidity, 10 - 85
  • atmospheric pressure, 91 kPa - 105.
Shipping method
Contact the seller
Compact analyzer "DAG-500"
Compact analyzer "DAG-500"
for
by the goods
Your message must contain at least 20 symbols. The message must not be more than 2000 symbols.
Obligatory field is not filled
The field is wrongly filled
The field is wrongly filled
Compare0
ClearSelected items: 0